Fluorescence Quantification

Fluorescence Colocalization in Fiji (Coloc 2 & JACoP): Pearson's r vs. Manders' M1/M2

Running Coloc 2 on a whole unmasked field of view almost always inflates Pearson's r because thousands of empty black background pixels agree that both channels are zero. Valid colocalization requires background subtraction, cellular ROI masking, and choosing the right metric for your biological hypothesis.

Segmented overlay for CTCF Fluorescence Intensity Quantification
Raw micrograph for CTCF Fluorescence Intensity Quantification
RawQuantified
Reference output for this workflowDrag the divider to compare the unmodified micrograph with automated quantification for CTCF Fluorescence Intensity Quantification.View CTCF Fluorescence Intensity Quantification specs

Pearson's r and Manders' M1 / M2 formulas

Pearson's correlation coefficient (r, range −1 to +1) tests whether intensities of Channel 1 and Channel 2 co-vary linearly pixel-by-pixel. Manders' split overlap coefficients (M1 and M2, range 0 to 1) report the fraction of total fluorescence in one channel that overlaps above-threshold signal in the second channel:

Formula
r=∑i(Ri−Rˉ)(Gi−Gˉ)∑i(Ri−Rˉ)2∑i(Gi−Gˉ)2,M1=∑i: Gi>TGRi∑iRir = \frac{\sum_{i}(R_i - \bar{R})(G_i - \bar{G})}{\sqrt{\sum_{i}(R_i - \bar{R})^2 \sum_{i}(G_i - \bar{G})^2}}, \qquad M_1 = \frac{\sum_{i:\,G_i > T_G} R_i}{\sum_{i} R_i}

R_i and G_i are background-subtracted pixel intensities in the two channels inside the cellular ROI mask; T_G is the Costes or fixed positive threshold for Channel 2.

Step-by-step Fiji Coloc 2 procedure

Always work on unmodified 16-bit single channels before merging any display colors:

  1. Split channels

    Open your raw multi-channel stack via Bio-Formats and run Image Color Split Channels.

  2. Subtract uneven background

    Run Process Subtract Background (Rolling Ball radius 50 px, check Sliding paraboloid on uneven fields) on each fluorescence channel before measuring Manders' coefficients.

  3. Mask out empty background

    Draw or segment cell/organelle ROIs and add them to Analyze Tools ROI Manager so empty black space outside cells is excluded from correlation.

  4. Launch Coloc 2

    Open Analyze Colocalization Coloc 2, assign Channel 1 and Channel 2, select your ROI/Mask, set PSF (typically 3 px for confocal), and set Costes randomisations to 100.

  5. Verify Costes p-value

    Confirm Costes P-value ≥ 0.95 (indicating correlation exceeds 95% of spatially randomized PSF blocks) and record both Pearson's r (no threshold) and tM1 / tM2.

Choosing between Pearson's r and Manders' M1 / M2

Pearson's r and Manders' M1/M2 answer distinct biological questions—reporting the wrong one leads to misleading conclusions when expression stoichiometry differs:

Biological questionRecommended metricWhy it fits
Do protein A and protein B co-vary stoichiometrically in the same compartment?Pearson's r (inside cell ROI)Sensitive to linear intensity proportionality; unaffected by uniform brightness scaling.
What fraction of protein A sits inside lysosomes (even if lysosomes outnumber protein A)?Manders' tM1 and tM2Separates A-in-B (M1) from B-in-A (M2) cleanly when one compartment is much larger.
Are discrete puncta overlapping above random chance?Object-based centroid / JACoP + Costes p-valueResists diffuse cytoplasmic haze that blurs pixel-level correlation.

Four pitfalls that invalidate colocalization data

Check optical and preprocessing controls before attributing high r or M1 values to real molecular interaction:

Key verification checkpoints
  • Spectral bleed-through:Always include single-stained control samples; if GFP leaks into the red detector, both Pearson's r and Manders' M1 will look artificially high.
  • Running Coloc 2 on full unmasked fields: Empty black background inflates Pearson's r as cell density drops across wells.
  • Skipping background subtraction before Manders': A uniform camera offset of 100 counts biases the denominator of M1/M2 unless subtracted consistently.
  • Chromatic aberration or z-drift:Verify alignment with multi-spectral calibration beads; even a 200 nm lateral shift sharply reduces pixel overlap on diffraction-limited puncta.